Chlorophyll Fluorescence vs SIF

8 min read

If you have ever clipped a fluorometer onto a leaf, waited for the flash and read off Fv/Fm, you have measured chlorophyll fluorescence. Solar-induced fluorescence (SIF) is the same light, emitted by the same chlorophyll molecules. The difference lies in how it is measured: a fluorometer brings its own light source and works on one leaf at a time, while SIF is measured with nothing but sunlight, from a distance, across a whole field. This article explains what the two methods share, where they differ, and which one fits which question.

SIFcam fluorescence image of a Monstera plant in sunlight: one leaf that was kept in the dark glows brightly just after being uncovered, the neighbouring leaf stays dim
Chlorophyll fluorescence at 760 nm, imaged with a SIFcam in plain sunlight from about nine metres away. The bright leaf had been covered and was uncovered a moment before; its neighbour was in the sun all along. No flash, no laser, no leaf clip. Recorded by Floronics, August 2026.

One signal, two ways to measure it

When chlorophyll absorbs light, the energy can go three ways: it drives photosynthesis, it is released as heat, or it is re-emitted as faint red and far-red light between about 650 and 850 nm [1][10]. That re-emitted light is chlorophyll fluorescence. Because the three pathways compete for the same energy, a change in photosynthesis or in heat dissipation shows up as a change in fluorescence [1][2]. This is why fluorescence has been a standard probe of photosynthesis in plant science for decades [1][4]. Our article on SIF walks through all five pathways that sunlight can take in a leaf.

The emission itself does not care where the absorbed light came from. A measurement, however, has to choose. Active methods illuminate the leaf with a controlled light source and record the response. Passive methods use the light that is already there, the sun, and record what the plant emits under real field conditions. Pulse-amplitude-modulation (PAM) fluorometry is the best-known active method. SIF is the passive one [5][10].

How a PAM fluorometer works

A PAM fluorometer uses a weak, modulated measuring light that excites chlorophyll without driving photosynthesis. On a dark-adapted leaf this gives the minimum fluorescence, Fo. A brief saturating flash of less than a second then closes all reaction centres of photosystem II, so that photochemistry can no longer take any energy, and fluorescence rises to its maximum, Fm [1][3]. From these two readings follows Fv/Fm, the maximum quantum yield of photosystem II. In unstressed leaves it is remarkably constant at about 0.83, and it falls when a plant is under stress, which makes it a simple and widely used stress indicator [1][3].

A field technician kneeling in a maize trial: one hand holds a leaf clip on a single maize leaf, a cable connects the clip to a portable fluorometer with a display
How an active fluorescence measurement looks in the field: the clip holds one leaf, the instrument supplies the light and reads the response. One leaf, one reading.

In the light, the same trick with the saturating flash separates the two processes that compete with fluorescence. It yields the operating efficiency of photosystem II (ΦPSII) and the level of protective heat dissipation, known as non-photochemical quenching (NPQ) [1][3]. That ability to split the energy budget is the great strength of the method.

It comes at a price. Fv/Fm and NPQ need a dark-adapted reference, which in practice means covering each leaf for 20 to 30 minutes [3]. And because the instrument has to control the light on the leaf, the method stays close to it: PAM fluorometry is mostly a single-leaf technique [5]. Handheld instruments have brought it from the lab bench into breeding nurseries and field trials [6], but the unit of measurement is still one spot on one leaf.

How SIF is measured

Under sunlight, every green leaf fluoresces all day long. Measuring that emission passively is hard for one reason: the fluorescence is buried in reflected sunlight that is far brighter. Passive instruments solve this by looking inside narrow dark lines of the solar spectrum, where the sunlight reaching the ground is strongly weakened and the plant's own emission stands out. Most often these are the oxygen absorption bands of the atmosphere, which are wide and deep enough to measure in and lie close to the fluorescence peaks. One of them is the oxygen A-band near 760 nm [5][7].

The result is a different kind of number. A PAM fluorometer reports yields and ratios. A SIF instrument reports how much fluorescence light leaves the canopy, as a radiance in physical units, under the light the plants are actually growing in [2]. Since no light source has to be carried to the leaf, the measurement works from a tower, a drone, an aircraft or a satellite, and an imaging instrument records every plant in view at the same moment.

Side by side

The table summarises the practical differences. Neither column is the better one. They describe two instruments built for two different scales.

Active fluorescence (PAM)SIF (passive)
Light sourceThe instrument's own measuring light and saturating flashesSunlight only
What you getYields and ratios: Fv/Fm, ΦPSII, NPQFluorescence radiance in physical units, as a map
Dark adaptationNeeded for Fv/Fm and NPQ, typically 20 to 30 minutesNone
Distance to the plantLeaf clip or a few centimetres; laser- or LED-based systems reach tens of metresMetres to hundreds of kilometres: tower, drone, aircraft, satellite
CoversOne spot on one leaf per readingEvery plant in the image at once
Ambient lightLargely independent of it, works at nightNeeds steady sunlight
Spectral windowA broad band, usually above 700 nmA very narrow band inside an absorption line
Strongest atExplaining the mechanism in selected leavesComparing many plots under the same conditions

Where the two meet: a textbook experiment in sunlight

The most famous experiment in fluorescence research is also the simplest. Keep a leaf in the dark, then expose it to light. Its fluorescence jumps to a maximum within about a second and then declines over seconds to minutes, as photochemistry and heat dissipation get going [1][2][3]. Kautsky and Hirsch described this in 1931 [8], and the effect, named after Kautsky, is still how most students first meet chlorophyll fluorescence, usually with a fluorometer in a dark lab.

We repeated it without a fluorometer. One leaf of a plant standing in full sun was covered for a while, its neighbour was not. A SIFcam about nine metres away filmed both while the cover was pulled off. The covered leaf flares up to about 3.5 mW m⁻² sr⁻¹ nm⁻¹ within about a second, more than three times the level of its neighbour. Five to six seconds later it has lost half of that surplus, and after two minutes it has settled at about 1.5. The neighbouring leaf stays at about 1.1 throughout.

The recording as a video, 30 seconds in real time. Each frame is a map of SIF at 760 nm; the colour scale runs from 0 to 6 mW m⁻² sr⁻¹ nm⁻¹.
012340306090120Seconds after the cover is removedSIF at 760 nm (mW m⁻² sr⁻¹ nm⁻¹)coveredLeaf kept in the darkNeighbouring leaf, in the sun all along
The Kautsky effect in plain sunlight: the same recording as the video above, as numbers. SIF at 760 nm of the two leaves (median over each leaf). The grey band marks the time before the cover was removed. A single recording of one plant, measured with a SIFcam from about nine metres away, August 2026.

The exact timing will differ from plant to plant and from day to day. The principle will not: the fast, physiological response that fluorometers are built to capture is also present in the passive signal. Simultaneous measurements on the same leaf area point the same way. PAM readings and spectrally resolved, passively recorded fluorescence were strongly correlated across the whole emission spectrum [9].

What SIF does not give you

SIF and Fv/Fm answer different questions. The saturating flash is what lets a fluorometer tell how much of the absorbed energy goes into photochemistry and how much is released as heat. SIF works without a flash, so it shows the result of both processes together [10]. In bright light, which is when field trials are flown, fluorescence and photosynthesis tend to rise and fall together. In dim light they can move in opposite directions [2][10]. In full sun SIF therefore follows photosynthesis, but it is not a fixed fraction of it.

A canopy is also not a leaf in a clip. How much fluorescence reaches the camera depends on how much light the canopy absorbs and on how much of the emitted light escapes from it. In rice, wheat and maize canopies these two, more than leaf physiology, explained how SIF related to photosynthesis over days and seasons [11]. In a trial this is handled by design and by processing: plots are compared within one flight under the same light, and the signal can be normalised by incoming light and by canopy reflectance. Pulling the physiological part out of the canopy signal ever more cleanly is where much of the current work in the field is going, including ours.

And SIF needs the sun. It is a daytime measurement that works best under a steady, clear sky, whereas a fluorometer can be used in a growth chamber at midnight.

Which tool for which question

Why not simply flash a whole field? Because an active measurement does not scale to one. Delivering a light pulse strong enough to saturate photosynthesis from a distance is technically demanding [5]. Active systems built for larger distances get around this with rapid sequences of short laser or LED pulses that only partly saturate the leaf, and work over distances of up to about 50 metres [5]. On an automated platform over crops grown in containers, such a system recorded genotype-specific photosynthetic responses in barley, maize, soybean and wheat over two growing seasons [12]. But it scans the canopy line by line, 30 cm at a time at about 3 cm per second [12]. That suits a row of containers or a handful of plots, not a trial with several hundred.

Take a breeder or trial manager as an example: the choice follows from the question. If you want to know why a particular line reacts the way it does, measure selected leaves with a fluorometer. If you want to know which of several hundred plots react at all, and when, you need a measurement that covers all of them within minutes and under the same light. That is where imaging SIF from a drone comes in. SIFcam images SIF at 760 nm with a ground resolution of 2.6 cm per pixel from 25 m. In a peer-reviewed drone campaign over a wheat and faba bean trial, SIFcam values agreed with a ground reference spectrometer at R² = 0.93 across 13 plots [13]. Field trials are only one example. Imaging SIF from a drone is still a young technique, and if you have another use in mind, we are glad to talk it through.

SIF map of a whole field trial seen from above: four rows of rectangular plots in colours from dark purple to bright yellow, separated by dark paths
A whole field trial in one SIFcam map, about 38 by 22 metres. Brighter means more fluorescence at 760 nm; paths and bare soil stay dark. Every plot is measured in the same flight, and the differences between neighbouring plots are visible at a glance.

The two methods work well together. A SIF map shows where in the trial something is happening. A fluorometer, taken to exactly those plots, helps explain what.

References
  1. 1.Baker, N. R. (2008). Chlorophyll Fluorescence: A Probe of Photosynthesis In Vivo. Annual Review of Plant Biology, 59, 89–113. https://doi.org/10.1146/annurev.arplant.59.032607.092759
  2. 2.Porcar-Castell, A., Tyystjärvi, E., Atherton, J., van der Tol, C., Flexas, J., Pfündel, E. E., et al. (2014). Linking chlorophyll a fluorescence to photosynthesis for remote sensing applications: mechanisms and challenges. Journal of Experimental Botany, 65(15), 4065–4095. https://doi.org/10.1093/jxb/eru191
  3. 3.Murchie, E. H., & Lawson, T. (2013). Chlorophyll fluorescence analysis: a guide to good practice and understanding some new applications. Journal of Experimental Botany, 64(13), 3983–3998. https://doi.org/10.1093/jxb/ert208
  4. 4.Maxwell, K., & Johnson, G. N. (2000). Chlorophyll fluorescence—a practical guide. Journal of Experimental Botany, 51(345), 659–668. https://doi.org/10.1093/jexbot/51.345.659
  5. 5.Mohammed, G. H., Colombo, R., Middleton, E. M., Rascher, U., van der Tol, C., Nedbal, L., et al. (2019). Remote sensing of solar-induced chlorophyll fluorescence (SIF) in vegetation: 50 years of progress. Remote Sensing of Environment, 231, 111177. https://doi.org/10.1016/j.rse.2019.04.030
  6. 6.Kuhlgert, S., Austic, G., Zegarac, R., Osei-Bonsu, I., Hoh, D., Chilvers, M. I., et al. (2016). MultispeQ Beta: a tool for large-scale plant phenotyping connected to the open PhotosynQ network. Royal Society Open Science, 3(10), 160592. https://doi.org/10.1098/rsos.160592
  7. 7.Meroni, M., Rossini, M., Guanter, L., Alonso, L., Rascher, U., Colombo, R., & Moreno, J. (2009). Remote sensing of solar-induced chlorophyll fluorescence: Review of methods and applications. Remote Sensing of Environment, 113(10), 2037–2051. https://doi.org/10.1016/j.rse.2009.05.003
  8. 8.Kautsky, H., & Hirsch, A. (1931). Neue Versuche zur Kohlensäureassimilation. Die Naturwissenschaften, 19(48), 964. https://doi.org/10.1007/BF01516164
  9. 9.Magney, T. S., Frankenberg, C., Fisher, J. B., Sun, Y., North, G. B., Davis, T. S., Kornfeld, A., & Siebke, K. (2017). Connecting active to passive fluorescence with photosynthesis: a method for evaluating remote sensing measurements of Chl fluorescence. New Phytologist, 215(4), 1594–1608. https://doi.org/10.1111/nph.14662
  10. 10.Porcar-Castell, A., Malenovský, Z., Magney, T., Van Wittenberghe, S., Fernández-Marín, B., Maignan, F., et al. (2021). Chlorophyll a fluorescence illuminates a path connecting plant molecular biology to Earth-system science. Nature Plants, 7(8), 998–1009. https://doi.org/10.1038/s41477-021-00980-4
  11. 11.Dechant, B., Ryu, Y., Badgley, G., Zeng, Y., Berry, J. A., Zhang, Y., et al. (2020). Canopy structure explains the relationship between photosynthesis and sun-induced chlorophyll fluorescence in crops. Remote Sensing of Environment, 241, 111733. https://doi.org/10.1016/j.rse.2020.111733
  12. 12.Keller, B., Matsubara, S., Rascher, U., Pieruschka, R., Steier, A., Kraska, T., & Muller, O. (2019). Genotype specific photosynthesis x environment interactions captured by automated fluorescence canopy scans over two fluctuating growing seasons. Frontiers in Plant Science, 10, 1482. https://doi.org/10.3389/fpls.2019.01482
  13. 13.Salattna, S., Elibol, A., Siegmann, B., Kneer, C., Scharr, H., Rascher, U., & Bendig, J. (2026). High Spatial Resolution Imaging of Solar-Induced Chlorophyll Fluorescence (SIF) From an Uncrewed Aerial Vehicle (UAV): Development and Evaluation of Image-Processing Workflows. IEEE Journal of Selected Topics in Applied Earth Observations and Remote Sensing, 19, 30350–30367. https://doi.org/10.1109/JSTARS.2026.3717082
Frequently asked
What is the difference between chlorophyll fluorescence and SIF?
+
SIF is chlorophyll fluorescence, measured passively. In both cases the light comes from chlorophyll in the photosynthetic machinery. A fluorometer excites it with its own lamp and reports yields such as Fv/Fm. SIF is the fluorescence a plant emits under natural sunlight, measured from a distance as a radiance in physical units. You may also see it called sun-induced fluorescence.
Can chlorophyll fluorescence be measured from a drone?
+
Yes, as solar-induced fluorescence. A drone cannot carry a saturating flash to every leaf of a field, but it does not have to: under sunlight the plants fluoresce on their own, and a SIF camera records that emission in a narrow band near 760 nm. SIFcam does this as an image, with a ground resolution of 2.6 cm per pixel from 25 m.
Do I need dark adaptation for SIF?
+
No. Dark adaptation is needed in PAM fluorometry to obtain Fv/Fm and NPQ. SIF is measured on plants in their normal, light-adapted state in the field, so there is nothing to cover and no waiting time.
Can SIF give me Fv/Fm or NPQ?
+
No. Both need a saturating light pulse, which a passive measurement does not have. SIF tells you how much fluorescence a canopy emits under the current light. It responds to changes in photosynthesis and in heat dissipation, but it cannot separate the two on its own.
Does SIF replace my fluorometer?
+
No, the two complement each other. A fluorometer explains the mechanism on the leaves you select. SIF compares all plots of a trial at the same moment and shows where it is worth looking closer. Leaf-level readings are also a useful ground reference when interpreting SIF maps.
Is the SIF signal as strong as what a fluorometer sees?
+
The emission is the same, but a passive instrument has to find it underneath reflected sunlight, where fluorescence makes up only about 1 to 5 percent of the light leaving a canopy at these wavelengths. That is why SIF is measured inside narrow absorption lines such as the oxygen A-band, and why it needs carefully calibrated optics.